BOSTON SCIENTIFIC CORPORATION EXPRESS LD ILIAC / BILIARY; STENT, ILIAC Back to Search Results
Model Number 20226
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was cancelled.A 8.0x40x135cm and a 7.0x60x135cm express-b-i ld stents were advanced for treatment.However, during the procedure, both stents failed to advance through sheath.The procedure was aborted and there were no patient complications reported.
 
Event or Problem Description
It was reported that the procedure was cancelled.A 8.0x40x135cm and a 7.0x60x135cm express-b-i ld stents were advanced for treatment.However, during the procedure, both stents failed to advance through sheath.The procedure was aborted and there were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: express-b-i ld was received for analysis.The device was received advanced through the customers 6fr sheath.The recommended introducer sheath size for this express ld device as per express specification is a 6fr.A visual inspection found the sheath to be damaged beginning at the hub and extending distally for approximately 15mm.A visual and microscopic examination also found the lining of the sheath to be wrapped around the tip of the express device and had been pulled distally out of the sheath together with the balloon catheter.The lining of the sheath was also found to be wrapped around the distal section of the device that had exited the sheath.A visual examination of the returned device confirmed that the balloon was tightly wrapped and had not been subjected to positive pressure.No issues were noted with the balloon material.A microscopic examination found the stent crimped in the correct location on the balloon.The first three rows of the distal stent struts were found to be lifted/damaged.This type of damage is consistent with excessive force being applied to the device.No issues were noted with the of the device's tip.The lining of the customer's sheath was noted to be tightly wrapped around the tip of the device.A visual and tactile examination identified no issues with the shaft of the device.No other issues were noted with the device that could potentially have contributed to the complaint incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12303633
Report Number2134265-2021-09852
Device Sequence Number19715946
Product Code NIO
UDI-Device Identifier08714729392453
UDI-Public08714729392453
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K133110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/14/2024
Device Model Number20226
Device Catalogue Number20226
Device Lot Number0026969523
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2021
Initial Date Received by Manufacturer 07/23/2021
Supplement Date Received by Manufacturer09/16/2021
Initial Report FDA Received Date08/10/2021
Supplement Report FDA Received Date09/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-