• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY Back to Search Results
Model Number PNML6F088904M
Medical Device Problem Codes Stretched (1601); Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a medical procedure in the middle cerebral artery (mca) to treat an arteriovenous malformation (avm) using a neuron max 6f 088 long sheath (neuron max).During the procedure, the physician experienced resistance while inserting a dilator into a neuron max and decided to remove the neuron max.Upon removal, it was noticed that the neuron max was stretched at the distal end and kinked at the proximal end.The procedure was completed using a non-penumbra sheath.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned neuron max confirmed that the catheter was stretched.If the neuron max is forcefully retracted against resistance, damage such as stretching may occur.Further evaluation revealed a kink in the proximal shaft.This damage was incidental to the reported complaint and the root cause could not be determined.During the functional test, the dilator was able to be advanced through the neuron max with resistance due to the kink and stretching on the catheter shaft.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON MAX 6F 088 LONG SHEATH
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12303684
Report Number3005168196-2021-01783
Device Sequence Number19826200
Product Code DQY
UDI-Device Identifier00814548012162
UDI-Public00814548012162
Combination Product (Y/N)Y
Initial Reporter CountryIN
PMA/510(K) Number
K111380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2023
Device Model NumberPNML6F088904M
Device Catalogue NumberPNML6F088904M
Device Lot NumberH10831
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/19/2021
Supplement Date Received by Manufacturer08/25/2021
Initial Report FDA Received Date08/10/2021
Supplement Report FDA Received Date09/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
Date Report Sent to FDA01/01/2005
-
-