AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL574T
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap ag that a challenger ti-p sm-ligat.Clips 12 cartr (part # pl574t) was used during a procedure performed on (b)(6) 2021.According to the complaint, the clip was jammed and came off.After loading the first shot, the clip was jammed and magazine came off and fell into the body.There was no described patient harm.Although requested, additional information has not been made available.The malfunction is filed under (b)(4) reference xc (b)(4).
 
Additional Manufacturer Narrative
Based upon investigation results, this event was re-evaluated and is considered no longer reportable due to risk file - no malfunction or serious injury.Investigation results: visual investigation: the investigation was carried out visually and microscopically.Here we detected a broken nose, clip jam and a deformed slider sheet.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 1(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Explanation and rationale: according to the quality standard and dhr files a material defect and a production error can be excluded.There are no hints of a pre- damage or something similar.Investigations lead to the assumption that the cartridge was engaged not completely or was damaged during inserting.If the cartridge is engaged not completely or was damaged during inserting, there is an impairment of product functionality.This could lead to deformed latches of the slider sheet, to clip jam and can cause the cartridge to jump off.Based upon historically grown product experience and due to different simulation regarding a cartridge which is engaged not completely, this leads to the described errors.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.
Common Device Name
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key12304659
Report Number9610612-2021-00537
Device Sequence Number3532451
Product Code FZP
UDI-Device Identifier04046963356057
UDI-Public4046963356057
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K081031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2026
Device Model NumberPL574T
Device Catalogue NumberPL574T
Device Lot Number52677252
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2021
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date08/10/2021
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-