| Model Number |
97725 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id: 977a260, serial#: (b)(4), product type: lead.Product id: 977a260, serial#: (b)(4), product: lead.Other relevant device(s) are: product id: 977a260, serial/lot #: (b)(4), ubd: 27-jan-2025, udi#: (b)(4).Product id: 977a260, serial/lot #: (b)(4), ubd: 03-dec-2024, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from the healthcare provider (hcp) via a manufacturer representative (rep) regarding a trial patient (pt) who was using wireless external neurostimulator (wens).It was reported that during intraop testing, we were getting multiple open circuit readings, and high impedance readings.Tried wiping the lead several times, and switching channels to determine if it was the lead or the wens with no success.The hcp instructed rep to swap out the wens.After switching to a new wens, the lead showed more open circuits and high impedance issues at multiple contacts, mainly the 4 most distal contacts.Switched out the lead with a new one.The issue was resolved.It¿s possible the lead was damaged when the hcp tried to bend the stylet to have more curvature. there is no patient involved in this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product id: 977a260, serial# (b)(6), product type: lead; product id: 977a260, serial# (b)(6), product type: lead.Analysis information -- 2021-09-08: the returned device passed all testing in the laboratory and no anomalies were identified.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the ins and leads indicated that there were no anomalies.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|