| Model Number |
16122-02-S |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device is currently being evaluated.A follow up report with results of the investigation will be submitted upon completion.
|
| |
|
Event or Problem Description
|
|
The targeting needle got stuck in sclerotic bone and broke which caused a delay in surgery.All pieces were removed from the patient.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: medical device problem code:1260 component code:752 type of investigation: 3331; 10 investigation findings:3252 investigations conclusions: 23 device evaluation: the device history record was reviewed.No irregularities were identified during manufacturing or inspection.The device returned on 07/16/2021.Visual inspection of the device confirmed the event.The top portion has completely de-attached from the device.The device was not returned for evaluation.A photograph was provided to confirm the event.The root cause is the result of the cannula hub interface yields/strips at a lower force than the force required to remove the product from bone.This is likely result related to the manufacturing process.
|
| |
|