• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXPRESS LD ILIAC / BILIARY; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXPRESS LD ILIAC / BILIARY; STENT, ILIAC Back to Search Results
Model Number 20226
Medical Device Problem Codes Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the stent moved on the balloon.The 80% stenosed target lesion was located in the non-tortuous and moderately severe calcified right iliac artery.A 7.0mm x 60mm x 135 cm express ld iliac stent was advanced for treatment without ballooning the lesion site.However, the device failed to cross the lesion even after several attempts and had to be removed.The lesion was dilated with a 5.0 balloon catheter, and then the stent was loaded on the wire again.After an attempt to deliver, the stent appeared loose on its balloon.The device was completely removed.The procedure was completed by ballooning the lesion site first and deploying another express ld stent.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12304945
Report Number2134265-2021-09831
Device Sequence Number3612108
Product Code NIO
UDI-Device Identifier08714729392453
UDI-Public08714729392453
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/03/2024
Device Model Number20226
Device Catalogue Number20226
Device Lot Number0027435859
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/22/2021
Initial Report FDA Received Date08/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-