|
Investigation: visual inspection: no significant deformations or other abnormalities of the valve were detected during the visual inspection.Permeability test: a permeability test has indicated that the valve has a blockage.Computer controlled test: to investigate the claim of over/under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.Because the valve is not permeable, a computer controlled test is not possible.Adjustability test: the test is not applicable, because the dsv is a fixed pressure valve.Results: first, we performed a visual inspection of the dsv.No significant deformations or damage of the valve were detected during the visual inspection.Next, we tested the permeability of the valve.The valve was shown to have a blockage.Then we carried out a computer controlled simulated flow test.Due to the nonpermeability of the valve, a computer controlled test was not possible to further investigate the claim of under-drainage.Finally, we have dismantled the valve.Inside the valve we have found a build-up of substances (likely protein).Based on our investigation, we confirm the presence of under drainage/occlusion in the valve, likely due to the deposits observed inside the valve.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|
|
It was reported that a dsv system w/prechamber 10/40dsv (part # fv173t) was implanted during a procedure performed on (b)(6) 2020.According to the complainant, the valve was believed to be operating in overdrainage and had a problem with adjustability.The patient underwent a revision procedure on (b)(6) 2021.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6) years, height: 193 centimeters (cm), weight: (b)(6) kilograms (kg), gender: male.
|