It was reported that the procedure was to treat a left anterior descending artery that was 70% stenosed.A 3x12mm xience alpine stent was advanced to the lesion without issue.The stent-balloon was pressurized at 10 atmospheres, but the balloon failed to inflate and therefore the stent failed to deploy.Contrast leakage was noted from the balloon area when pressurized.The device was removed and troubleshooting was performed at the prep table, but the stent-balloon continued to fail to inflate and leakage was noted from the balloon area.Another xience alpine stent was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
The device was returned for analysis.The reported material rupture was not confirmed.The reported activation failure was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported material rupture as a material rupture was not confirmed during return evaluation.It should be noted; however, a tear in the outer member was noted during return evaluation.It is likely that the outer member tear was misidentified as a balloon rupture and caused the reported activation failure.Factors that may contribute to outer member tears include, but are not limited to, material damage and interactions with other devices.There was no damage or leak noted to the stent delivery system (sds) during the inspection prior to use or during preparation of the device, which suggests a product quality issue did not contribute to the reported difficulties.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
|