This product is manufactured by zimmer biomet (b)(4) and is not cleared or distributed in the u.S.However, this report is being submitted as zimmer biomet (b)(4) manufactures a similar device that is cleared or distributed in the united states.The manufacturer did not receive devices or x-rays for review.Lot numbers were received for the device and the device history records were reviewed and found to be conforming.Based on an extensive investigation of events reported from several user facilities outside the usa, zimmer identified that the most probable cause for the outcome observed was a loose or unstable cup that resulted from use of surgical techniques not consistent with the manufacturer's recommendations.As a corrective action, a retraining program for users outside the usa was initiated in november 2009 and reported to the national competent authorities as required.The durom cup reported in this case is not marketed in the usa.A similar cup, compatible with the metasul ldh femoral head, is cleared in the usa and a corrective action for this product was reported to the fda in july 2008 as notification z-2415/2426-2008.Since this case is related to the issues for which zimmer implemented a corrective action there will be no further investigation.Zimmer¿s reference number of this file is (b)(4).
|