| Model Number |
120614 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the way tiemann catheters were packaged allowed them to become twisted, and the guide at the end of the coude tip was not positioned correctly, which caused the catheters not to be inserted correctly.
|
| |
|
Event or Problem Description
|
|
It was reported that the way tiemann catheters were packaged allowed them to become twisted, and the guide at the end of the coude tip was not positioned correctly, which caused the catheters not to be inserted correctly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be due to "inappropriate package design".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "indications for use: for urological use only.Warnings: on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra instructions for intermittent catheters: wash your hands thoroughly with soap and water.Remove catheter from the pack.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.Gently insert rounded end of catheter into urethra.When urine stops flowing, remove catheter from urethra.Dispose of catheter in accordance with local rules and regulations.Wash your hands." h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
|
| |
|