| Model Number |
FLT-100 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pulmonary Edema (2020)
|
| Date of Event |
07/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.The device was received and evaluated by pmqa after being received and decontaminated through the rma process.Visual inspection was complete, the iv pole and blue handle were found to be loose.Initial power on, no issues were noted.A linearity test was performed.Both load cells remained within the allowable tolerance.A deficit accuracy test was performed.No problems warnings or errors were observed.The reported complaint was not confirmed.Hologic will continue to track and trend this event.Complaint could not be confirmed/reproduced, unable to establish relationship between the issue reported and the device.We will continue to trend and monitor the reported condition.
|
| |
|
Event or Problem Description
|
|
It was reported that during a myomectomy that involved a fluent on july 16th, during the procedure the deficit jumped to 5000 and the staff had to perform a manual count.The procedure was completed and at the end of the procedure the manual count was 2500 ml.The patient had to be readmitted to the hospital after the surgery due to a pulmonary edema.No other information is available.
|
| |
|