MITG - RF SURGICAL SYSTEMS SITUATE; GAUZE/SPONGE, INTERNAL, X-RAY DETECTABLE Back to Search Results
Model Number G0404-16P01CN-1
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during procedure, the x-ray sponge was linting/shredding.Came off on patient's skin.There was no patient injury.
 
Event or Problem Description
According to the reporter, during breast surgery removal of axillary mas, the x-ray sponge was linting/shredding.Came off on patient's skin.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had an excessive fabric detached from gauze.Functional testing found the gauze showed fraying and linting.The gauze had been manipulated.It was reported that the product was linting.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: the gauze sponges are made with natural cotton and may produce lint or fibers when unfolded or manipulated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITUATE
Common Device Name
GAUZE/SPONGE, INTERNAL, X-RAY DETECTABLE
Manufacturer (Section D)
MITG - RF SURGICAL SYSTEMS
2101 faraday
carlsbad CA 92008
MDR Report Key12309842
Report Number3005883396-2021-00021
Device Sequence Number5002393
Product Code GDY
UDI-Device Identifier10884521700819
UDI-Public10884521700819
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG0404-16P01CN-1
Device Catalogue NumberG0404-16P01CN-1
Device Lot Number200517AT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2021
Initial Date Received by Manufacturer 07/27/2021
Supplement Date Received by Manufacturer10/12/2021
Initial Report FDA Received Date08/11/2021
Supplement Report FDA Received Date10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-