| Model Number |
DHT3/4 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet (b)(4).Patient identifier, age and date of birth.Gender, patient weight unknown / not provided.Date of event unknown / not provided.Pma/510(k) number unknown / not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that clinician indicated implant box was empty, it had only product information inside.No other product available in the dr.Office.Clinician was unable to finish the procedure as the box was empty.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay corrected data, additional information and device evaluation.The following sections are being reported: h6: component code was updated/corrected: 4747.H6: type of investigation codes were added: 4109, 4110, 4111, 4114 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 67.The xfr hex direct scr-v 3.5m m neck 4.5mm imp (dht3/4) and packaging were not returned, therefore, visual evaluation of the product could not be performed.The investigation was performed using applicable instructions for use and risk files to address reported event.Dhr review for the lot (2019040794) revealed no deviations nor non-conformances which could have caused or contributed to the reported event.All products were conforming at the time they left zimmer biomet.Lot was inspected and passed all acceptance criteria by qa.The dhr was further reviewed to verify the packaging process.No malfunction or nonconformance was identified.Complaint history review was performed for the reported lot number (2019040794) for similar events (complaint category keyword: packaging: incorrect component quantity, packaging other) and no other complaint was identified.November post market trending was reviewed and there were no actionable events or corrective actions for the reported event or products.Therefore, based on the available information, device malfunction and the reported event could not be verified since the condition of the products/packaging as received by the customer was unknown/non-verifiable.
|
| |
|
Event or Problem Description
|
|
No additional event information at the time of this report.
|
| |
|
Search Alerts/Recalls
|