• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 9/10 COCR HEAD 28MM +3; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 9/10 COCR HEAD 28MM +3; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1521-90-059
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/16/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2002, the patient underwent tha.After the surgery, at the regular medical examination this year (the date is unknown), partial wear of liner was confirmed by x-ray.On (b)(6) 2021, the patient underwent revision surgery.In the revision surgery, loosening at the neck and head of the stem was confirmed.Before the revision surgery, the replacement of the stem was not planned, so only the head, liner, and locking ring were replaced as scheduled, and the surgery was completed successfully without any surgical delay.When the neck of the stem was observed, there was a slight scratch and a small amount of black powder on the neck taper of the stem.It was confirmed by x-rays taken after the revision surgery that there was no problem.The surgeon is following up the patient.No further information is available.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9/10 COCR HEAD 28MM +3
Common Device Name
SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12310394
Report Number1818910-2021-17639
Device Sequence Number3962816
Product Code JDI
UDI-Device Identifier10603295276647
UDI-Public10603295276647
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K871867
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2002
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1521-90-059
Device Catalogue Number152190059
Device Lot NumberYKL-75
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/16/2021
Supplement Date Received by Manufacturer08/25/2021
Initial Report FDA Received Date08/11/2021
Supplement Report FDA Received Date08/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
9/10 COCR HEAD 28MM +3.; CANCELLOUS BONE SCREW 6.5X30MM.; CANCELLOUS BONE SCREW 6.5X35MM.; CANCELLOUS BONE SCREW 6.5X35MM.; DURALOC 1200 SERIES 54MM OD.; ENDURON 10D 54OD X 28ID.; REPLICA A STEM 13.0MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight64
-
-