BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MARTIN-LEWIS AGAR; CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA Back to Search Results
Model Number 221558
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using bd bbl¿ martin-lewis agar atypical growth was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "it was reported that there are "poor results.".
 
Additional Manufacturer Narrative
H.6.Investigation: during manufacturing of material 221558, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.No batch number, returns or photos were provided for investigation.Retention samples cannot be evaluated without a specific batch number.Trending was performed on the appropriate quality databases.A recall for three batches for martin lewis agar was issued on march 17, 2021 for misformulation: material 221558 batches 0317559 and 0345695 and material 221557 batch 0317558.No information provided from the customer acknowledges the use of the recall batches.The latest expiry of the recall batches was april 06, 2021 and this complaint was taken on july 22, 2021.This complaint cannot be confirmed.Bd will continue to trend complaints for performance.
 
Event or Problem Description
It was reported that while using bd bbl¿ martin-lewis agar atypical growth was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: " it was reported that there are "poor results".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ MARTIN-LEWIS AGAR
Common Device Name
CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12310853
Report Number1119779-2021-01362
Device Sequence Number3817868
Product Code JTY
UDI-Device Identifier10382902215582
UDI-Public10382902215582
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K770287
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number221558
Device Catalogue Number221558
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2021
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date08/11/2021
Supplement Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-