OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC PROBE Back to Search Results
Model Number UM-S20-17S
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc confirmed the reported phenomenon.It was found that there was a hole on the sheath at the distal end and fluid was entered inside the device.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of this event could not be conclusively determined.Omsc surmised that the black scratch was that the fluid which was entered from the hole fixed inside the sheath.The hole on the sheath could occur due to the following external force applied.The device was forcibly inserted or withdrawn into/from the endoscope with the bending section bent.During probe rotation, the device was forcibly inserted or withdrawn into/from the endoscope.
 
Event Description
Olympus medical systems corp.(omsc) was informed by the user that during the reprocessing, the user found something like a black scratch attached to the distal end of the subject device.The user reprocessed the subject device, but it could not be removed.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12311280
MDR Text Key266162508
Report Number8010047-2021-10130
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Was the Report Sent to FDA? No
Date Manufacturer Received07/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-