LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-01-04
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).Pictures of the reported error message were provided to livanova.A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.The drive unit was replaced with another.Investigation is ongoing.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump 5 (cp5) gave a runaway fault error message during priming and the user could not clear the alarm.The drive unit was replaced with another to initiate the procedure.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: the drive unit was requested for investigation.No deviation could be found during functional tests and the device worked within specifications.Based on the investigation performed for similar cases, the reported error message is most likely due to: hand cranking performed by the user while the cp5 control panel is still on; malfunction of the motor control board.No hand cranking was performed by the customer since it was reported that the drive unit was replaced and another drive unit was used to continue the procedure.Based on gathered data and investigation results, a systematic failure of the motor control board can be excluded.Thus, the reported event was most likely due to an intermittent failure of the motor control board which will be replaced as per equipment maintenance intervention prevention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key12312383
MDR Text Key266199190
Report Number9611109-2021-00446
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-04
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/02/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-