Catalog Number 530.705 |
Device Problems
Complete Loss of Power (4015); Intermittent Energy Output (4025)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the battery reamer device had insufficient/low power, intermittent operation and would not run.It was further determined that the device failed pretest for check forward/reverse mode function and check power with the test bench.It was noted in the service order that the device did not work at all.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
|
|