EIT EMERGING IMPLANT TECHNOLOGIES GMBH INSERTER S-HAMMER CONNECTION; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
| Model Number |
PET30102 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date the two instruments broke at the welding seam intraoperatively.No fragments were generated.The surgery was completed successfully without any delay and with another available instrument.No adverse patient harm.This report is for one (1) inserter s-hammer connection this is report 1 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: the device lot number is unknown, therefore a device history review could not be performed.If more information become available, the record will be re-assessed.Photo investigation.The device was not returned.A photo-investigation was performed on the images.Upon inspecting the image provided under notes & attachments section, the inserter sh connection was observed to be broken at the proximal end of the head of the shaft.The root cause for the reported event cannot be determined from the available information.The device lot number is unknown, therefore a mre review could not be performed.If more information become available, the record will be re-assessed as the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Conclusion.The complaint condition can be confirmed during photo/video investigation.During the investigation, no product design issues or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: a product investigation was conducted.Visual inspection: the inserter sh connection (p/n: pet30102, lot #: e19di1071) was received at us cq.Upon visual inspection of the device, it was observed that the pet30102 b plif inserter pin was broken at the laser weld next to the thread.No other issues were observed with the returned device.Dimensional inspection: the dimensional inspection of the device could not be conducted due to post-manufacturing damage.Document/specification review.Based on the date of manufacture, the current and manufactured revision of drawings were reviewed.Investigation conclusion.The complaint condition was confirmed for the received device.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.The potential cause could be due to unintended forces applied to the device.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: a device history record (dhr) review was conducted: part # e19di1071.Synthes lot # pet30102.Supplier lot # pet30102.Release to warehouse date: supplier: (b)(4) 18 nov 2019.No ncr's were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|