TELEFLEX MEDICAL VISISTAT 35W NON-STERILE; STAPLE, REMOVABLE (SKIN) Back to Search Results
Model Number IPN914814
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).The device history review for the product visistat 35w non-sterile lot# 73j2000213 investigation did not show issues related to the complaint.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
Customer complaints that stapler does not work properly - the staples penetrate only one side of the skin.The patient's current condition is unknown.The staples do not hold to the skin and a loose suture is seen with scar disunion.The risks for the patient were an immediate or later disunion leading to a healing and aesthetic defect.Local care with directed healing had to be implemented at home and there may be a need for remote revision surgery for cosmetic damage.The incident was resolved by using a medtronic product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISISTAT 35W NON-STERILE
Common Device Name
STAPLE, REMOVABLE (SKIN)
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key12313102
Report Number3003898360-2021-00682
Device Sequence Number3164766
Product Code GDT
UDI-Device Identifier04026704600779
UDI-Public04026704600779
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN914814
Device Catalogue Number528236
Device Lot Number73J2000213
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2021
Initial Report FDA Received Date08/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-