| Model Number |
1011496-12 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.The investigation was determined the reported failure to advance and stent dislodgement appear to be related to operational circumstances of the procedure.During advancement the device encountered resistance with the lesion causing the failure to advance.Manipulation against resistance likely resulted in the reported stent dislodgment during retraction.Additionally, the treatment appears to be related to the operational context of the procedure as the stent was removed using a snare device.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that this was a procedure to treat a lesion with heavy calcification in the left renal artery.The lesion was pre-dilated successfully.A 5.5 x 12 mm herculink stent delivery system (sds) was advanced but could not cross the heavily calcified lesion and during the attempt the stent dislodged.The dislodged stent remained on the guide wire.The sds was removed without issue.A snare device was used to help retrieve the dislodged stent.A new herculink sds was used to complete the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|