BOSTON SCIENTIFIC CORPORATION REZUM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number G2200
Medical Device Problem Codes No Display/Image (1183); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during preparation for two separate water vapor therapy procedure cases, the screen on the generator turned blue during priming of the delivery device.The staff rebooted the generator and were able to perform a few treatments.However, the generator screen turned blue again.The generator was turned off, unplugged from the wall and all connections were checked.The issue did not resolve, and in both instances, the procedure was aborted.There were no complications to the patient.This event is being reported for aborted/cancelled procedure with a patient under anesthesia.This report is for 1 of 2 procedures in which the generator was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REZUM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key12313754
Report Number2124215-2021-24421
Device Sequence Number4412017
Product Code GEI
UDI-Device Identifier00855357006003
UDI-Public00855357006003
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K180237
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG2200
Device Catalogue NumberG2200
Device Lot Number0000004336
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2021
Initial Report FDA Received Date08/12/2021
Patient Sequence Number1
-
-