MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 16026545
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product was returned for evaluation although a lateral radiograph was provided confirming the device malfunction.The patient was asymptomatic, but a revision occurred on a unknown date where the broken screws were completely removed and replaced with larger diameter screws.The patient is doing well post revision.The patients post op activity levels as well or whether the patient experienced a fall is unknown.It is unknown whether fusion occurred.Review of the provided information was inconclusive and the root cause cannot be determined at this time however the root cause is possibly related to excessive loading or post-operative physical activity.No additional investigation can be completed."potential adverse events and complications, as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s)." "correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants." "these devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patients activity level will dictate the longevity of the implant." "patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed.".
 
Event or Problem Description
On (b)(6) 2020 a female patient, of larger size, underwent a transforaminal lumbar interbody fusion procedure from the l5-s1 levels.On a unknown date it was discovered the patient experienced bilateral fractures of the implants screws at the s1 level.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
MDR Report Key12314000
Report Number2031966-2021-00089
Device Sequence Number3839164
Product Code NKB
UDI-Device Identifier00887517556639
UDI-Public887517556639
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K182974
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number16026545
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2021
Initial Report FDA Received Date08/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-