• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERGLIDE PRO MIDLINE CATHETER; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERGLIDE PRO MIDLINE CATHETER; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number F118081
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of reex1227 showed no similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "after deploying catheter into vein, wire sheared and remained in patient.It appears wire unraveled when attempting to remove.Removed wire from patient.Advised rn/md.Obtained xray of arm to determine if any wire remained in patient.".
 
Event or Problem Description
It was reported "after deploying catheter into vein, wire sheared and remained in patient.It appears wire unraveled when attempting to remove.Removed wire from patient.Advised rn/md.Obtained xray of arm to determine if any wire remained in patient.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a frayed guidewire is confirmed but the exact cause remains unknown.One 18 ga powerglide pro device was returned for evaluation.The product sticker label indicated lot: reex1227.The catheter was not returned with the device and the safety mechanism was successfully activated over the needle tip.Blood residue and evidence of use was present on the device.The guidewire was observed to be frayed and elongated.Microscopic observation of the core wire and coil wire did not reveal the weld tip to be present.The event description indicates fraying of the wire occurred during removal.Based on the information provided, retraction against resistance and damage against the needle bevel.Since the guidewire was observed to be frayed, the complaint is confirmed but the exact causing factors during use remain unknown.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERGLIDE PRO MIDLINE CATHETER
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key12314116
Report Number3006260740-2021-03261
Device Sequence Number17270078
Product Code FOZ
UDI-Device Identifier00801741110559
UDI-Public(01)00801741110559
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K153280
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberF118081
Device Catalogue NumberF118081
Device Lot NumberREEX1227
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/22/2021
Supplement Date Received by Manufacturer11/18/2021
Initial Report FDA Received Date08/12/2021
Supplement Report FDA Received Date11/24/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-