Model Number PM100N
Device Problems Device Alarm System (1012); Failure to Power Up (1476)
Patient Problem Low Oxygen Saturation (2477)
Event Date 07/14/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during use, the neonatal patient had a saturation reading of 52 for a considerable amount of time but the device did not alarm.Also, the unit had some physical damage to the casing, specifically, the battery compartment.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the keyboard/keypad/switch panel membrane was damaged on the monitor.Additionally, the main circuit board of the monitor was defective.It was reported that the product had an issue with the alarm system and there was an issue with the battery contacts.The reported issue was confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Type of Device
OXIMETER
Manufacturer (Section D)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KS  00000
MDR Report Key12314186
MDR Text Key266246200
Report Number2936999-2021-00787
Device Sequence Number1
Product Code DQA
UDI-Device Identifier10884521196728
UDI-Public10884521196728
Combination Product (y/n)N
PMA/PMN Number
K142865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 10/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM100N
Device Catalogue NumberPM100N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2021
Date Manufacturer Received09/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-