• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS HI FLEX ISRT TR S1-2 9; TRAY, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS HI FLEX ISRT TR S1-2 9; TRAY, SURGICAL, INSTRUMENT Back to Search Results
Model Number 71430401
Medical Device Problem Codes Crack (1135); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that one (1) genesis ii posterior stabilized high flexion insert trial size 1-2 9 mm, one (1) echelon 260 mm porous bowed femoral component standard collar size 16 left, one (1) genesis ii posterior stabilized high flexion insert trial size 7-8 9 mm, one (1) genesis ii posterior stabilized high flexion insert trial size 7-8 11 mm, one (1) genesis ii posterior stabilized high flexion insert trial size 5-6 9 mm and one (1) genesis ii posterior stabilized high flexion insert trial size 3-4 11 mm are cracked and worn.As this was noticed upon inspection, there was no patient involvement.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT TR S1-2 9
Common Device Name
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12314231
Report Number1020279-2021-06314
Device Sequence Number17103588
Product Code FSM
UDI-Device Identifier03596010493019
UDI-Public03596010493019
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71430401
Device Catalogue Number71430401
Device Lot Number05AM19777
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/28/2021
Supplement Date Received by Manufacturer10/08/2021
Initial Report FDA Received Date08/12/2021
Supplement Report FDA Received Date10/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-