Model Number PM3562
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the pacemaker exhibited incorrect measurement.The mode switch percentage increased to 37% since the device was last cleared when there was only one episode lasting 6 seconds since the time of clearing.It was because there was an ongoing episode at the time of the clearing which did not count towards the diagnostics until after the episode ended.Patient was asymptomatic and will continue with regular follow up.
 
Manufacturer Narrative
The complaint of an unexpected increase in alternate mode switch (ams) percentage was confirmed.The cause of the increase was a result of management of data related to an ams episode that was ongoing during the first session and ended prior to the second.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12315030
MDR Text Key266272125
Report Number2017865-2021-28074
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2021
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberA000087996
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/16/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-