NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE PLATING SYSTEM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
|
Back to Search Results |
|
| Model Number |
LP4.5-PLF4B |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
Osteolysis (2377)
|
| Date of Event |
07/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Information was received that upon explant, corrosion was noted around the plate's male/female junction.At the time of the removal bridging bone was present, however, osteolysis and scalloping of the regenerate bone occurred in close proximity to the corroded plate junction in the femur.
|
| |
|
Search Alerts/Recalls
|
|
|