• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI; OXIMETER Back to Search Results
Model Number WW1000EN
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a sensor issue was indicated.
 
Additional Manufacturer Narrative
One unit was returned for investigation.The device was hooked up to a live patient and had no readings which confirmed the reported complaint issue.The issue was found to be the spo2 probe which is a supplied item.No further information will be available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI
Common Device Name
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
MDR Report Key12315860
Report Number3012307300-2021-08380
Device Sequence Number14383397
Product Code DQA
UDI-Device Identifier30843418000016
UDI-Public30843418000016
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberWW1000EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/13/2021
Supplement Date Received by Manufacturer09/20/2021
Initial Report FDA Received Date08/12/2021
Supplement Report FDA Received Date10/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-