• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9734680
Device Problem Material Integrity Problem (2978)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 07/20/2021
Event Type  malfunction  
Manufacturer Narrative
No parts have been returned to the manufacturer for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding spinal instruments during a l1-s1 revision and fusion.It was reported that the doctor was using a mallet to advance the thoracic probe through the pedicle and into the vertebral body for screw preparation.After malleting the probe, the doctor removed it and noticed the tip of the probe broke off at the 10mm mark.The doctor used the passive planar probe and placed it down the hole the tip was located in.The doctor concluded the tip of the broken probe was too far into the vertebral body to retrieve it without causing harm to the patient.It was noted that around 10mm of the tip of the thoracic probe was left in the patient as the physician concluded they would cause more harm trying to retrieve it then if they left it.The relationship between the reported event and the use of medtronic products was unknown.The patient status was alive with no injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key12316419
MDR Text Key266391637
Report Number1723170-2021-02077
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number9734680
Device Catalogue Number9734680
Device Lot Number180719
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/20/2021
Initial Date FDA Received08/12/2021
Supplement Dates Manufacturer Received09/22/2021
Supplement Dates FDA Received09/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age69 YR
Patient Weight103
-
-