| Model Number |
381211 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that 2 bd insyte-n¿ iv catheters 24ga 0.56in experienced a needle that pierced through the needle cover.The following information was provided by the initial reporter: nurses and registrar noted on preparing for iv cannular insertion, bd insyte-n 24 ga 14mm, that the needles were going through the side of the plastic sleeve, also one we opened had a right angle bend at its tip.Registrar went to insert iv cannular into 3month old baby, but the guiding needle had gone out the side of plastic sleeve (part that is left in vein) this prickled skin but could not be advanced so failed and had to re-cannulate.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: no photos or samples were received by our quality team for evaluation therefore the failure mode could not be verified, and the root cause could not be determined.A device history record could not be evaluated as the lot number is unknown.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported that 2 bd insyte-n¿ iv catheters 24ga 0.56in experienced a needle that pierced through the needle cover.The following information was provided by the initial reporter: nurses and registrar noted on preparing for iv cannular insertion, bd insyte-n 24 ga 14mm, that the needles were going through the side of the plastic sleeve, also one we opened had a right angle bend at its tip.Registrar went to insert iv cannular into 3month old baby, but the guiding needle had gone out the side of plastic sleeve (part that is left in vein) this prickled skin but could not be advanced so failed and had to re-cannulate.
|
| |
|