| Model Number |
TLEAD1058-50B |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Fever (1858); Unspecified Infection (1930); Pain (1994)
|
| Date of Event |
07/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The manufacturing records were reviewed and no relevant nonconformities were found.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported to nevro that the patient experienced pocket pain and a fever.Nevro attempted to obtain additional information regarding the nature of the issue but none was available.The device was removed and there have been no reports of further complications regarding this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is to update based on new information.
|
| |
|
Event or Problem Description
|
|
Follow-up indicated that the patient's pocket pain and fever were due to an infection caused by a bandage changed at home.The physician drained the infection and there have been no reports of further complications regarding this event.
|
| |
|