OLYMPUS WINTER & IBE GMBH WORKING INSERT, ALBARRAN, TWO WAY; WORKING INSERTS WITH ALBARAN LEVER Back to Search Results
Model Number A20972A
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Perforation (2001)
Date of Event 07/16/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.Please note: as of today the customer did not provide the model number for the effected device.Therefore, we entered a probable number.
 
Event or Problem Description
Olympus was informed that during an unspecified therapeutic procedure the albarran perforated the patient's urethra causing a tear in several places of the urothelium.The albarran was twisted 90 degrees.No further information was provided.
 
Additional Manufacturer Narrative
The suspect medical device was not returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient¿s outcome and the reported phenomenon could not be conclusively determined and is being judged as unknown.However, based on the information provided, the cause for the patient injury is very likely improper handling by the customer.It is assumed that the user removed the working insert (with the erect albarran) out of the patient¿s body together with the sheath and the optical instrument, instead of pulling it out through the sheath, which serves as place holder in the patient¿s body.A dhr review could not be performed since basic data of article identification (lot number) are missing.Instead, a manufacturing and quality control review was performed for the last 24 months of production.There were no non-conformities or deviations regarding the described issue.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.Please note: as of today the customer did not provide the model number for the effected device.Therefore, we entered a probable model number.
 
Additional Manufacturer Narrative
Additional information: d4 ¿ lot number; h4 ¿ device manufacturer date the suspect medical device was returned to olympus for evaluation/investigation.During the investigation both the pull wires and the albarran lever were found to be missing.Furthermore, the distal end of the telescope channel is severely deformed and shows visible tool marks.The proximal end of the telescope channel shows dents and further deformations and it is no longer possible to insert a telescope.It is very unlikely that this severe damage occurred during use, rather it is assumed that it was caused afterwards by improper handling by the customer.Based on the information provided, it is assumed that the albarran and its pull wires were deformed due to mechanical overload.The cause for the reported patient injury is very likely improper handling by the customer.It is assumed that the user removed the working insert (with the erect albarran) out of the patient¿s body together with the sheath and the optical instrument, instead of pulling it out through the sheath, which serves as place holder in the patient¿s body.A manufacturing and quality control review was performed for the affected lot number of the working insert without showing any abnormalities.The case will be closed from olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKING INSERT, ALBARRAN, TWO WAY
Common Device Name
WORKING INSERTS WITH ALBARAN LEVER
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key12317746
Report Number9610773-2021-00226
Device Sequence Number14154766
Product Code GCP
UDI-Device Identifier04042761020374
UDI-Public04042761020374
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA20972A
Device Catalogue NumberA20972A
Device Lot Number19703-0011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/30/2021
Supplement Date Received by Manufacturer09/02/2021
10/26/2021
Initial Report FDA Received Date08/13/2021
Supplement Report FDA Received Date09/03/2021
10/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-