| Model Number |
SAM350P |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Heartsine has requested return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
|
| |
|
Event or Problem Description
|
|
Inability to use the device shock button, may prevent or delay defibrillation which could result in adverse consequences.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Heartsine evaluated the customer's device and verified the event was logged in the device's memory; however, the reported issue could not be duplicated in testing, and the investigation was inconclusive as no fault was found on the returned device.The device was scrapped by heartsine and the customer was provided with a replacement device.
|
| |
|
Event or Problem Description
|
|
The distributor contacted heartsine to report that their device was showing the "shock key stuck" error message.Inability to use the device shock button, may prevent or delay defibrillation which could result in adverse consequences.There was no patient involvement reported with this event.
|
| |
|