• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER ACCESS SARS-COV-2 IGG II; IN VITRO DIAGNOSTIC DEVICES: CORONAVIRUS SEROLOGICAL REAGENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER ACCESS SARS-COV-2 IGG II; IN VITRO DIAGNOSTIC DEVICES: CORONAVIRUS SEROLOGICAL REAGENTS Back to Search Results
Catalog Number C69057
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The customer did not provide patient demographics such as age, date of birth, gender, weight, ethnicity or race.The access sars-cov-2 igg ii assay was not returned for evaluation.There were no reports of system issues at the time of the event.There was no report of issues with other assays at the time of the event.No hardware errors or flags were reported in conjunction with the event.The patient sample was submitted to the beckman coulter complaint handling unit (chu) for testing.The chu obtained one non-reactive and one reactive result as the result was close to the cut-off of the assay, close to the results obtained by the customer.The chu then tested the sample with the sars-cov-2 igg 1st is assay.Non-reactive results were generated, suggesting the sample is non-reactive for sars-cov-2.Sars-cov-2 is an enveloped non-segmented positive-sense rna virus.It has several structural proteins including spike (s), envelope (e), membrane (m) and nucleocapsid (n).The spike protein (s) contains a receptor binding domain (rbd) which is responsible for recognizing the cell surface receptor, angiotensin converting enzyme-2 (ace2).It is found that the rbd of the sars-cov-2 s protein strongly interacts with the human ace2 receptor leading to endocytosis into the host cells and viral replication.The access assay detect antibodies directed against the spike protein, which are more likely to neutralize the virus.Different vaccines were placed on the market and do not elicit the same immune response.The roche assay is a total assay, detecting igg, igm and iga without distinction.The results of a total assay cannot be compared to the results of an ¿igg-only¿ assay.Additionally, the access sars-cov-2 igg ii assay is not currently labelled for vaccine response detection and is not currently standardized against the international standard.The performances of the access sars-cov-2 igg ii or access sars-cov-2 igg (1st is) assays have not been established in individuals that have received a covid-19 vaccine.While a positive antibody test result can be used to help identify people who may have had a prior sars-cov-2 infection, more research is needed in people who have received a covid-19 vaccination.In conclusion, the cause of the event cannot be determined with the available information.
 
Event or Problem Description
On (b)(6) 2021 the customer reported non-reactive covid igg results (access sars-cov-2 igg ii, part number c69057, lot number 124558) were generated on the customer's dxi (dxi 800 access immunoassay analyzer, part number 973100 and serial number (b)(4)).The customer did not indicate whether the results were released from the laboratory.The customer did not report a change to patient care of treatment in connection with this event.The customer reported that the patient had been vaccinated against sars-cov-2 with the astrazeneca vaccine in early june.Customer reported the patient had a reactive covid result with a sample collected (b)(6) 2021 and tested on the roche anti-sars-cov-2 s total method on (b)(6) 2021.The customer also reported a sample collected and tested on (b)(6) 2021 and tested using the access covid igg assay had also had a nonreactive result.No hardware errors were reported in conjunction with this event.System performance indicators such as system check, calibration and quality control were passing within specifications ad the time of the event.No issues with sample integrity were reported by the customer.Sample was noted to be collected in a 5 ml bd vacuette serum tube with clot activators.It was noted to be a full draw and allowed to clot for >30 minutes; centrifuged sample was noted to be clear.Centrifugation time and speed not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS SARS-COV-2 IGG II
Common Device Name
IN VITRO DIAGNOSTIC DEVICES: CORONAVIRUS SEROLOGICAL REAGENTS
Manufacturer (Section D)
BECKMAN COULTER
1000 lake hazeltine drive
chaska MN 55318
MDR Report Key12318689
Report Number2122870-2021-00128
Device Sequence Number2825837
Product Code QKO
UDI-Device Identifier15099590742744
UDI-Public(01)15099590742744(17)211031(11)210503(10)124558
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Catalogue NumberC69057
Device Lot Number124558
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2021
Initial Report FDA Received Date08/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-