Model Number 1504-10-209
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Synovitis (2094); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 12/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.Health effect - clinical code: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
(b)(6) 2015: patient underwent a right tka due to djd.Patella was resurfaced and depuy cement was used.No complications noted.(b)(6) 2019: patient underwent a right tka revision due to aseptic loosening and instability.Intraoperative findings included - swelling/effusion, medial laxity, fibrous scarring and adhesions, and blood tinged synovial fluid (synovitis) that was encroaching on the joint causing impingement.There was a fragmentation of the cement along peripheral margins of the femoral component.Femoral component was noted to be well fixed, but slightly oversized.The tibial insert showed wearing and pitting on medial side.The patella had some flat spot with osteolytic changes.The tibial component was grossly loose with separation of the component from the cement.Medial tibial bone loss.All depuy implants were removed and replaced with competitor implants.(b)(6) 2020: patient underwent an insert exchanged due to infection - competitor products involved.Doi: (b)(6) 2015.Dor: (b)(6) 2019.Right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM RT SZ 9 CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key12319490
Report Number1818910-2021-17826
Device Sequence Number4441438
Product Code JWH
UDI-Device Identifier10603295041801
UDI-Public10603295041801
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number1504-10-209
Device Catalogue Number150410209
Device Lot Number7962725
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/23/2021
Supplement Date Received by Manufacturer08/24/2021
10/20/2021
Initial Report FDA Received Date08/13/2021
Supplement Report FDA Received Date08/25/2021
10/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 9 CEM; ATTUNE MEDIAL DOME PAT 41MM; ATTUNE PS FB INSRT SZ 9 5MM; ATTUNE PS FEM RT SZ 9 CEM; SMARTSET GHV GENTAMICIN 40G; SMARTSET GHV GENTAMICIN 40G; ATTUNE FB TIB BASE SZ 9 CEM; ATTUNE MEDIAL DOME PAT 41MM; ATTUNE PS FB INSRT SZ 9 5MM; ATTUNE PS FEM RT SZ 9 CEM; SMARTSET GHV GENTAMICIN 40G; SMARTSET GHV GENTAMICIN 40G
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-