| Model Number |
384157 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is expected to return for evaluation.A follow-up report will be submitted once the sample has been received and reviewed.
|
| |
|
Event or Problem Description
|
|
Picc implantation was performed for the patient on june 22, and fluid leakage was found on july 31.
|
| |
|
Additional Manufacturer Narrative
|
|
One catheter was returned for review and leakage was confirmed just below the inverted triangle beneath the securement disc.Careful handling is required during the procedure as not to cause damage to the soft silicone catheter.The ifu warns users: do not use hemostats or clamps on catheter or hub connection.Never use catheter for high-pressure injection.Syringes smaller than 10 ml and mechanical high-pressure injectors can generate pressures capable of rupturing the catheter.Never use force to flush the catheter if resistance is met.Solutions containing high concentrations of alcohol can temporarily weaken the structure of polyurethane material.For polyurethane catheters, minimize exposure to alcohol solutions.Do not expose the catheter to acetone or acetone/alcohol solutions.Do not use if package is opened or damaged.Do not cut stylet.If cut, stylet can potentially damage the catheter and harm the patient.Never use force to advance the catheter.Resistance could indicate a vein obstruction or malposition of the catheter.Never use force to remove the stylet.Resistance can damage the integrity of the catheter and the stylet.Do not hold the catheter with forceps while removing the stylet.Syringes smaller than 10 ml can generate high pressures (greater than 40 psi) capable of rupturing the catheter.Never place tape strips over the catheter tubing.This will compromise the strength and integrity of the tubing.
|
| |
|
Event or Problem Description
|
|
Picc implantation was performed for the patient on (b)(6), and fluid leakage was found on (b)(6).
|
| |
|
Search Alerts/Recalls
|