• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. LP SYSTEM DETACHMENT HANDLE; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. LP SYSTEM DETACHMENT HANDLE; HCG, KRD Back to Search Results
Model Number RLPH1
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.This report is associated with mfr report numbers: 3005168196-2021-01810, 3005168196-2021-01811, 3005168196-2021-01812.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat an arteriovenous malformation (avm) using packing coil lps, a lp system detachment handle (handle) and a non-penumbra microcatheter.During the procedure, the physician attempted to detach three packing coil lps using the handle; however, all three packing coil lps failed to detach.The physician also tried using another handle but was unsuccessful.Afterwards, the physician was able to manually detach the packing coil lps after multiple attempts.It was reported that the first handle worked fine while performing a click test; however, while trying to detach the coil, the handle would only half click and slowly return to its original position without detaching the coil.The procedure was completed using five additional packing coil lps and the second handle.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LP SYSTEM DETACHMENT HANDLE
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12321330
Report Number3005168196-2021-01813
Device Sequence Number4136109
Product Code HCG
UDI-Device Identifier00815948021969
UDI-Public00815948021969
Combination Product (Y/N)Y
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K192955
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/13/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRLPH1
Device Catalogue NumberRLPH1
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2021
Initial Report FDA Received Date08/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-