Model Number 1219-36-058
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient received a corial / pinnacle metal on poly thr 8 years ago, 12 months ago reported of hearing a squeaking noise, extra 12 months ago showed no issues.However a recent x-ray 6 weeks ago showed that the poly liner had flipped out of the cup.Upon retrieval one of the ard¿s on the liner has snapped off and the liner was positioned m/l in the cup instead of a/p.Signs of wear on the liner where the metal head had been skimming off between the liner and directly into the cup.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : a worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated the following: doi: appx 8 years ago, dor: (b)(6) 2021, affected side: right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 36IDX58OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12324978
Report Number1818910-2021-17910
Device Sequence Number4002370
Product Code LPH
UDI-Device Identifier10603295014478
UDI-Public10603295014478
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K010171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1219-36-058
Device Catalogue Number121936058
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/27/2021
Supplement Date Received by Manufacturer08/18/2021
09/22/2021
Initial Report FDA Received Date08/16/2021
Supplement Report FDA Received Date08/25/2021
09/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M HEAD 36MM +12; CUP; PINN MAR NEUT 36IDX58OD; ARTICULEZE M HEAD 36MM +12; CUP; PINN MAR NEUT 36IDX58OD
Outcome Attributed to Adverse Event Required Intervention;
-
-