DEPUY ORTHOPAEDICS INC US PINNACLE 300 ACET CUP 56MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-03-056
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Pain (1994); Osteolysis (2377); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Medical records from (b)(6) 2021 note that the patient has severe left hip pain after a stumble and fall.The implants consisted of a corail uncemented femoral stem, and a pinnacle cup with a metal on metal bearing surface.Doi: (b)(6) 2010, dor: (b)(6) 2021.Left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : a worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 300 ACET CUP 56MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12325345
Report Number1818910-2021-17920
Device Sequence Number17144407
Product Code LPH
UDI-Device Identifier10603295008675
UDI-Public10603295008675
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1217-03-056
Device Catalogue Number121703056
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/20/2021
Supplement Date Received by Manufacturer10/11/2021
Initial Report FDA Received Date08/16/2021
Supplement Report FDA Received Date10/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
12/14 ARTICUL 40MM M SPEC-2; CORAIL HIGH OFFSET STEM SZ 10; PINNACLE 300 ACET CUP 56MM; PINNACLE MTL INS NEUT40IDX56OD; 12/14 ARTICUL 40MM M SPEC-2; CORAIL HIGH OFFSET STEM SZ 10; PINNACLE 300 ACET CUP 56MM; PINNACLE MTL INS NEUT40IDX56OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient Weight87
-
-