| Model Number |
27803 |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported that the emma provided high values.No consequences or impact to patient were reported.
|
| |
|
Event or Problem Description
|
|
The customer reported that the emma provided high values.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated and passed functional testing.During accuracy testing, the unit measured outside of the specification.A zeroing calibration was performed and the following measured value was within accuracy specification.Incorrect zeroing results in co2 measurements outside the accuracy specification.Performing zeroing restored accuracy.A service history record review reveals that this unit was in the field for over eleven (11) months with no previous reported issues related to this reported event.
|
| |
|