• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LOCKING SCREW, PARTIALLY THREADED T2 TIBIA 5X50 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LOCKING SCREW, PARTIALLY THREADED T2 TIBIA 5X50 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1891-5050S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 07/23/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
 
Event or Problem Description
It was reported the devices 1891-5050s and 1986-5050s were broken due to non-union, but were successfully removed along with other retained hardware.Revision with hoffmann lrf external fixation.
 
Event or Problem Description
It was reported the devices 1891-5050s and 1986-5050s were broken due to non-union, but were successfully removed along with other retained hardware.Revision with hoffmann lrf external fixation.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.A device inspection was not possible since the affected device was not returned, and no other evidences were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.General aspects: screw breakage in general has been experienced.It does not present an unanticipated event in itself.Depending on the load application, also, depending on the patient¿s post implant behaviour, on the suitable anatomical reduction, on the kind of bone breakage and depending on the course of bone healing and other factors a screw breakage can rather be classified as anticipated ¿ specifically if one or more contributing issues concurrence with each other.A successful treatment with an implant requires sufficient bone healing during the implantation period.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, PARTIALLY THREADED T2 TIBIA 5X50 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
MDR Report Key12329016
Report Number0009610622-2021-00639
Device Sequence Number14630879
Product Code HSB
UDI-Device Identifier04546540201959
UDI-Public04546540201959
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K032244
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/31/2024
Device Model Number1891-5050S
Device Catalogue Number18915050S
Device Lot NumberK040559
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/23/2021
Supplement Date Received by Manufacturer09/30/2021
Initial Report FDA Received Date08/16/2021
Supplement Report FDA Received Date10/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-