• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION 1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK 1.5MM X 5MM; SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION 1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK 1.5MM X 5MM; SCREW Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Concomitant medical product - unknown blade catalog#: ni lot#: ni, unknown driver catalog#: ni, lot#: ni.Report source - foreign - (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated report:0001032347-2021-00399.
 
Event or Problem Description
It was reported that the screw fractured at the head during an insertion attempt.The surgeon was able to remove the screw body from the patient with the use of a drill.No consequences or impact to the patient.It was reported that no further information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d4; g3; g6; h1; h2; h3; h6; h10 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Review of complaint history identified additional similar complaints for the reported items and no additional complaints for the reported part and lot combination.Medical records were not provided.A definitive root cause cannot be determined.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK 1.5MM X 5MM
Common Device Name
SCREW
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key12329095
Report Number0001032347-2021-00401
Device Sequence Number14141719
Product Code JEY
UDI-Device Identifier00841036053741
UDI-Public00841036053741
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K121589
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number91-6105
Device Lot Number845730
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/22/2021
Supplement Date Received by Manufacturer08/31/2021
Initial Report FDA Received Date08/16/2021
Supplement Report FDA Received Date09/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-