• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 M.R.I. ULTRA SLIMPORT IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 M.R.I. ULTRA SLIMPORT IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 0605640
Device Problems Contamination (1120); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 11/2024).
 
Event Description
It was reported that prior to a procedure, foreign material was allegedly found on the device.There was no patient contact.
 
Event Description
It was reported that prior to the procedure, foreign material was allegedly found on the device.There was no patient contact.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed bard access system port kit was returned for evaluation.Gross visual and microscopic evaluation were performed.The investigation is confirmed for the reported contamination issue as cardboard-like fibrous foreign material was noted within the top clear window of the outer package.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 11/2024).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that prior to the procedure, foreign material was allegedly found on the device.There was no patient contact.
 
Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed bard access system port kit was returned for evaluation.Gross visual and microscopic evaluation were performed.The investigation is confirmed for the reported contamination issue as cardboard-like fibrous foreign material was noted within the top clear window of the outer package.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.R.I. ULTRA SLIMPORT IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12329663
MDR Text Key266837210
Report Number3006260740-2021-03343
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741025990
UDI-Public(01)00801741025990
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K924250
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0605640
Device Catalogue Number0605640
Device Lot NumberREEV2624
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-