• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM/130 DEG TI CANN TFNA 235MM/RIGHT - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM/130 DEG TI CANN TFNA 235MM/RIGHT - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 04.037.044S
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: the investigation could not be completed; no conclusion could be drawn, as no product was received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, a damaged trochar and fns fell on the floor during an unknown procedure.There was no patient involvement.This report is for one (1) 10mm/130 deg ti cann tfna 235mm/right - sterile.This is report 1 of 1 for (b)(4).
 
Event or Problem Description
This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Manufacturing location: monument.Manufacturing date: 16-jul-2020.Expiration date: 31-may-2030.Part number: 04.037.044s, 10mm/130 deg ti cann tfna 235mm/right ¿ sterile.Lot number: 61p3013 (sterile).Lot quantity: 6.Production order traveler met all inspection acceptance criteria.Inspection sheet, in process / inspect dimensional / final, met all inspection acceptance criteria.Inspection sheet, tfna assembly inspection, met all inspection acceptance criteria.Packaging label logs (pll) lppf were reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿implants fell on the floor¿ does not indicate breakage of the nail or any of its components.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Visual inspection: the 10/130 deg ti cann tfna 235/right (part.No: 04.037.044s, lot.No: 61p3013, qty: 1) was returned and received at us customer quality (cq).Visual inspection of the complaint device showed no damage.Device failure/defect identified? no.Complaint confirmed? no.Investigation conclusion: this complaint is not confirmed as no defects were identified with the complaint device.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM/130 DEG TI CANN TFNA 235MM/RIGHT - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12330820
Report Number2939274-2021-04693
Device Sequence Number16818317
Product Code HSB
UDI-Device Identifier10886982096005
UDI-Public(01)10886982096005
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.037.044S
Device Catalogue Number04.037.044S
Device Lot Number61P3013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2021
Initial Date Received by Manufacturer 08/06/2021
Supplement Date Received by Manufacturer08/18/2021
Initial Report FDA Received Date08/17/2021
Supplement Report FDA Received Date08/31/2021
Patient Sequence Number1
-
-