• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRIO MESH; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRIO MESH; SURGICAL MESH Back to Search Results
Catalog Number 0010213
Medical Device Problem Codes Defective Device (2588); Patient-Device Incompatibility (2682); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Adhesion(s) (1695); Hernia (2240); Insufficient Information (4580)
Date of Event 11/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventrio patch on (b)(6) 2012.As reported, the patient is making a claim for an adverse patient outcome against the ventrio patch.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-may-2012) is considered to be a best estimate.Updated fields: a2, b3, b4, b5, b7, d4 (product catalog no., corporate lot no., udi no., expiry date), d.6b (date of explant), g3, g6, h2, h4, h6, h10.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventrio patch on (b)(6) 2012.As reported, the patient is making a claim for an adverse patient outcome against the ventrio patch.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2012 - patient was diagnosed with incisional hernia thereby underwent open repair with the implant of ventrio mesh.Per operative notes, ¿hernia sac was dissected out.A ventrio mesh was placed into the preperitoneal space and secure it with sutures.¿ (b)(6) 2020 - patient was diagnosed with recurrent umbilical hernia, adhesion thereby underwent laparoscopic repair with explant of ventrio mesh and implant one synthetic mesh.Per operative notes, ¿hernia sac was dissected out.Previously placed ventrio mesh was removed.All adhesions were taken down.A synthetic mesh was placed and secure it with sutures.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRIO MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key12331876
Report Number1213643-2021-06920
Device Sequence Number19539432
Product Code FTL
UDI-Device Identifier00801741016400
UDI-Public(01)00801741016400(17)150528(10)HUWE1005
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K100229
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/28/2015
Device Catalogue Number0010213
Device Lot NumberHUWE1005
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/13/2021
Supplement Date Received by Manufacturer08/21/2025
Initial Report FDA Received Date08/17/2021
Supplement Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/15/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Patient SexFemale
-
-