| Model Number |
DRV,3.5MM HEX,CANN (MINI HUDSON) |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during calcaneus medialisierunc surgery the tip did break off while screwing in the screw.The broken off piece has been retrieved from the patient.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.No further information received.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, the drive feature broke off.Available critical dimensions were found to meet specifications.No other abnormalities were observed.The cause of the event is undetermined, however likely causes are user-applied mechanical forces; prying/leveraging the device.
|
| |
|