| Model Number |
414-159 |
| Medical Device Problem Codes |
Break (1069); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Unintended Radiation Exposure (4565)
|
| Date of Event |
07/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was discarded, thus no investigation could be completed.
|
| |
|
Event or Problem Description
|
|
A peripheral atherectomy procedure commenced to treat a slightly calcified lesion in the patient's mid superficial femoral artery (sfa).The physician chose to use a spectranetics turbo elite laser atherectomy catheter to treat the patient.After the physician used the turbo elite device in the patient and was removing the device, he noticed the "outer sheath was torn" and fibers were visible.It was reported that there was difficulty getting the turbo elite device through the sheath and believed that may have been the reason the device was damaged.A new spectranetics turbo elite laser atherectomy catheter was used to finish the procedure, and there was no reported patient harm.This event is being submitted to capture the turbo elite which was found damaged during use and resulted in unintended radiation exposure with potential for harm.
|
| |
|