MAUDE Adverse Event Report: COOPERSURGICAL, INC. LEEP SYSTEM 1000 ESU GEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOPERSURGICAL, INC. LEEP SYSTEM 1000 ESU GEN Back to Search Results
Model Number 52969
Device Problem Mechanical Problem (1384)
Patient Problem Insufficient Information (4580)
Event Date 07/13/2021
Event Type  malfunction  
Manufacturer Narrative
Coopersurgical, inc.Is currently investigating the reported condition.
 
Event Description
Foot p.Not working.Order: (b)(4).Replaced diagram.(b)(4).
 
Event Description
Foot p.Not working.Order: (b)(4).Replaced diagram.1216677-2021-00165: leep system 1000 esu gen 52969 e-complaint: (b)(4).
 
Manufacturer Narrative
Investigation: x-inspect returned samples.Analysis and findings: complaint (b)(4).Distribution history: this complaint unit was manufactured at csi on 10/16/2009 under wo #(b)(4) and shipped on (b)(6) 2009.Manufacturing record review: a review of the device history record could not be located at the time of this investigation.However, it should be noted at the time of manufacture records from each lot are thoroughly reviewed.Should the device history record be located going forward this complaint will be amended accordingly.Incoming inspection review: not applicable.Service history record: no additional service history records found for this unit.Historical complaint review: a review of the 2-year complaint history showed similar reported complaint condition.Product receipt: the complaint unit was returned on repair log 96682.Visual evaluation: visual examination of the complaint unit revealed no physical damage.Functional evaluation: complaint unit was functionally evaluated and found not to function properly.Root cause: the foot pedal is used to activate the device.If not functioning the power to the unit will not flow.The foot pedal has been known to fail due to a few reasons.The most common is the diaphragm in the switch which acts as the mechanism to make contact (for electrical current) to supply power to the unit.Air displacement pushes on a piston via the diaphragm in the switch to turn the power on.The foot pedal creates the air displacement to the pneumatic switch.Given the appearance of a bad diaphragm it appears to have been exposed to excessive stimuli.Latex is flexible hence the use in this application but its elastic properties can alter over time as well.In this condition, the sealing properties are impacted.The root cause for this complaint condition is component related to the diaphragm.Correction and/or corrective action.Preventative action activity: the unit was repaired and returned to the customer.Sustaining engineering has successfully tested a replacement material made of silicone for use in assembly and repairs going forward.The ifu was also updated to add a safety check via ecn-20444, p/n 35387b.A service bulletin was issued to existing customers informing them to check for this issue and return the unit if needed.All product in fg and sk, as applicable, were reworked to replace the previous versions of the dfus on all applicable products.Correction activity in 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEEP SYSTEM 1000 ESU GEN
Type of Device
LEEP SYSTEM 1000 ESU GEN
Manufacturer (Section D)
COOPERSURGICAL, INC.
95 corporate drive
trumbull CT 06611
Manufacturer (Section G)
COOPERSURGICAL, INC.
75 corporate drive
trumbull CT 06611
Manufacturer Contact
michael marone
50 corporate drive
trumbull, CT 06611
4752651665
MDR Report Key12333228
MDR Text Key268451868
Report Number1216677-2021-00165
Device Sequence Number1
Product Code HGI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K952483
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number52969
Device Catalogue Number52969
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/20/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-