| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self-test for defib and pacer functions.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation.The customer's report was observed during initial testing.However, after disassembling the device to determine root cause, the report was no longer seen.The processor/bridge/pace board was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|